Journal: Brain Stimulation
Authors: Roth Y, Pell GS, Ankry M, Hadad Y, Eisen A, Burnishev Y, Tendler A, Moses E, Zangen A
Background:
The effect of unidirectional TMS is limited to neurons that are parallel to the induced electric field. With rotational field TMS (rfTMS), two orthogonal coils are operated with a 90° phase shift between them. The electric field vector is circularly polarized and rotates affecting neurons in various orientations.
Objective:
To compare the physiological effects of deep-rotational-field–TMS(rfdTMS) coils to unidirectional-dTMS(udTMS).
Methods:
Two orthogonal H7-coils were connected to a dual-channel stimulator to control timing and polarity. Hand and leg rMT were measured with EMG for eight healthy volunteers using: a. rfdTMS with four polarity states of the two coils, in each state a different 270° portion of the phase space is spanned by the field vector. b. udTMS by operating each coil separately, or simultaneously with no phase shift, in all polarity combinations in eight orientations 45° apart. Additionally, we measured for each condition the MEP and latency evoked at 120% rMT with the single H7 coil oriented posteroanterior for the hand and leg.
Results:
rMT with rfdTMS was significantly lower than any udTMS orientation for both hand (73.1±2.4%) and leg (73.2±0.7%). Suprathreshold MEPs were significantly higher with rfdTMS for both the hand (551±194%) and the leg (840±334%) compared to the single H7 coil oriented posteroanterior. For the hand, an angle of 45° to posteroanterior axis yielded the highest MEP (275%±54%), while for the leg there was no significant difference. Hand MEP latencies were not significantly different between rfdTMS and udTMS, while for leg, rfdTMS resulted in significantly shorter latencies.
Conclusions:
rfdTMS led to lower rMT and higher supra-threshold MEP, compared to udTMS. The efficiency of the rotational field effect is increased at depth where the fields overlap compared to superficial levels of the cortex that primarily experience the field from one of the coils.
Full Publication: Comparing rotational-field-dTMS to unidirectional-dTMS in healthy volunteers – Brain Stimulation: Basic, Translational, and Clinical Research in Neuromodulation